Click words or phrases color for definitions and descriptions
When a word or phrase is the title of a chapter, a link is provided to give direct access to the chapter
It is not possible to individually control each medical device being reprocessed. Instead, each reprocessing step (from point of use processing to aseptic presentation) undergoes validation to minimize the risk of a non-sterile product.
Initial validation takes place before a new process or equipment is implemented. Revalidation ensures that process effectiveness is maintained over time. It takes place periodically (often annually) or after a major event (for example, voluntary modification or breakdown)
Between validations (initial validation and revalidation) periodic or systematic routine controls preserve process effectiveness over time.
The validation plan defines the operations, methods and schedules of the initial validations, revalidation and routine controls as well as the responsibilities, training and, as needed, accreditations of the personal in charge.
Validation includes the qualification of the equipments used by the process. Qualifications of equipment consist of installation qualification (IQ) , operational qualification (OQ) ,performance qualifications (PQ) and requalifications (rQ).
In the present guidelines it will be considered that all processes, including those predominently manual, require equipment (for example a bench with sink possibly equipped with pumps for manual cleaning) and will hence consist of IQ, OQ, PQ. Validation of manual processes such as point of use processing or manual cleaning is key to minimize human dependance. Risks of deviance is however inevitably be higher than for a validated automated process
Installation qualification (IQ) is intended to confirm that the services associated with the equipment (water, electricity, etc.) and any accessories and safety devices are installed in accordance with the manufacturer’s instructions.
Documentary verification may be considered as a prerequisit to qualification operations or as being part of IQ.
The documents to be made available by the manufacturer vary according to the type of equipment but have some common points. Before installation, the supplier:
In return, the future user specifies any installation constraints (eg quality of the water available), the requirements in terms of the test method to be used.
After installation, the user checks:
Language in which the user and service manual must be made available is defined by local regulation. |
Operational qualification (OQ) is carried after successful IQ. It is intended to verify that the equipment performs according to manufacturer design specifications and international standards respectively. OQ tests are performed with empty equipment or test loads.
Additionally, the calibration status of all instruments (including test instruments) used for monitoring, control, indication or recording should be confirmed.
Performance qualification (PQ) takes place after successful IQ and OQ on an equipment in normal operating conditions. All factors that may have an influence (operators, methods, materials and the environment) should be taken into account.
Requalification is usually performed annually periodically or earlier in case of an event that may affect the efficacy of the process.
Examples of events justifying a requalification: • Change in equipment and installation that may impact process performance (e.g. modification of cycle parameters, new load carriers or cleaning products). • New load, unless it has been shown to be equivalent to another load already validated. • Failure detected during a routine control • Relocation of the equipment or technical intervention with change of critical elements in the PLC (or intervention on control elements) |
The contents of the initial IQ, OQ and PQ and scheduled requalifications are predefined. Scheduled requalifications may be simplified compared to initial qualifications (for example IQ will not be repeated).
The content of the unscheduled validation is event related (for example IQ may be partially or totally repeated after a relocation of the equipment).
Validation and requalifications operations are supervised by the sterilization department or the technical department of the healthcare facility. The healthcare facility may however contract with the supplier of the equipment or third parties. A detailed written description of the scope, objectives and respective assignments must be made produced.
In practice IQ is performed by the supplier of the equipment (manufacturer or distributor) or a service provider accredited by the supplier. Some or all OQ operations may be performed by the user, by a qualified third party or by the supplier as part of a service contract (or a service provider accredited by the supplier) PQ may be performed by the user, by a third party or by the supplier (or a service provider accredited by the supplier) Routine controls are usually carried out by the user in accordance with his documented procedures |
International standards do not assign responsibilities but they say that the responsibilities and party accepting the responsibilities must be defined and recorded.
Validation and requalification reports are approved and kept for record by the healthcare facility.They are grouped in a process validation report confirming:
for equipment, the validation (or revalidation) report includes
|
The validation of Automated Endoscope Reprocessor will be added in a future update of the guidelines. Washer-disinfector includes a disinfection phase, which in some case
The validation of Automated washer-Disinfector includes the validation of a disinfection stage (thermal or chemical). For some devices the disinfection stage might be the last step before use
Some local guidelines say that validation of Washer-Disinfector complying to ISO 15883-6 and ISO 15883-7 can be simplified as they pertain to non-critical items
IQ verifies that the connections and utilities (power, water, drain, dosing systems etc..) and Washer disinfector, ultrasonic bath or manual cleaning workstation are installed according to manufacturer specifications, that the equipment is safe for use and that equipment and consumables documentation is available.
OQ verifies that the equipment operates as specified by the manufacturer (for example, absences of leaks, chemistry dosing, temperature for thermal washer-disinfector, cavitation for ultrasonic baths, drying). Cleaning test are performed with test soils.
Controls are visual for visible surface and quantitative (e.g. protein tests) for complex device with hidden surfaces. Test methods, test soils and acceptance criteria are described in ISO 15883-58 (see below)
PQ is intended to control that reproducible results are obtained in conditions fully representative of real life. Control tests are performed with actual loads or devices contaminated by clinical use. Validated surrogate devices and test soils may however be used when real devices are difficult to sample.
Prior to PQ, load or devices are grouped in families presenting common processing characteristics. Load or device used for PQ are selected to present the highest cleaning challenge for a given family.
Controls are visual for external surface, carriers and loads and quantitative for complex devices with hidden surfaces. Test methods, test soils and acceptance criteria are described in ISO 15883-5,8 (see below)
The 2005 version of ISO 15883-5 listed soils and test methods developped by different countries. ISO 15883-5 ( 2021), lists chemical substance (analytes) contained in artificial or clinical soils.
![]() |
![]() |
Annex B of ISO 15883-1 explains the A0 concept.
A0 is the equivalent value in seconds at 80 ° C to produce a given disinfection effect on a specified microorganism. When the microorganism is a bacterial spore (the most resistant) we speak of A0. A0 = ∑ 10 ((T - 80) / Z) ∆t T is the temperature Z is specific to each microorganism. The Z value indicates the increase in T ° C allowing to accelerate the reduction of the population by a factor of 10 (in other words a change in the D of the microorganism by a factor of 10). For bacterial spores the z is 10 ° C. ∆ t is the holding time at a temperature ≥ T Standard ISO 15883-1 imposes a minimum A0 of 600 s corresponding to 10 min at 80 ° C or 1 min at 90 ° C. It adds that any washer-disinfector must offer at least one cycle capable of achieving an A0 of at least 3000 s (i.e. 5 minutes at 90 ° C, 2 mins 30 at 93 ° C). The recommendations of the standard on the positioning of sensors are detailed below: |
ISO 15883-1 is being adapted according to ISO 15883-5 and recent evolution. As experience is acquired with the new standard framework, adaptations may be proposed by local guidelines
An example of Manual cleaning processes validation inspired by ISO 15883-1 and ISO 15883-5 principles may be as follows
The “guideline for validation of manual cleaning and manual chemical of medical devices” 9 published in Germany is an example of a validation plan for manual cleaning processes
![]() |
Validation of ultrasonic cleaning bath must be adapted to the use of the type and objective of the ultrasonic cleaning process.
Cavitations test can be performed using:
|
An example of ultrasonic cleaning processes validation inspired by ISO 15883-1 and ISO 15883-5 principles may be as follows
![]() |
Packaging validations is governed by standards ISO 11607-110 (packaging system), 11607-211 (packaging process) and guided by ISO 1677512
ISO 11607-2 concerns packaging processes such as forming, sealing and assembly of preformed sterile barrier systems, sterile barrier systems and packaging systems. ISO 16775 provides guidance on ISO 11607-1 and -2 and suggests validations plans and checklists for each type of packaging (wraps, pouches and reels and containers).
The packaging validation plan lists the types of packaging and packaging equipment and determines, for each type of packaging (wraps, pouches, reels, containers), those that will be subject to validation.
The installation qualification (IQ) controls that:
The user is not expected to repeat the type tests already performed by the manufacturer. Instead, certificates of compliance to the tests listed in the appendix to NF EN ISO 11607-1 and recommended by the applicable NF EN 868 standard should be requested (867-213 for wraps, 867-514 for Pouches and reels, 867-813 for containers). The chemical indicators affixed to the packaging should comply with NF EN ISO 11140-1.
The operational qualification (OQ verifies the correct implementation of various packaging, controls that heat sealers meet manufacturer specifications and control containers are operational and were maintained according to plans. PQ test are performed with the most challenging load for each type of packaging.
For wraps: controls of correction execution
. |
The performance qualification controls the entire packaging circuit from preparation to aseptic presentation (article 6.3 ISO 11607-1 January 2018)). Performance validation is performed with performed with actual content representing the most challenging packaging and handling, transport and storage conditions. It confirms, through practice, the risk analyzes carried out when defining the packaging system suitable for each RMD or family of DMRRMD’s.
Within each family of packaging (sheets, sachets / sheaths, containers) several variants, specific to each hospital, may exist (according to the type of DMR, sterilization methods, size, brand etc…) . In practice, it is accepted to limit operational and performance validations after apropriate risk analysis. ISO 16775 provide guidance to determine the size of OQ samples. ISO 16775 suggests 3 evaluations for each type of packaging (i.e. 3 loads and 3 sterilization cycles. One of them corresponds to the conditions considered to be the most penalizing. If destructive checks are carried out (dye test or peel test), the number of samples tested is increased accordingly.
Requalification is annual. It may be simplified compared to the initial validation. A risk analysis will determine if a requalification is required after an incident or a voluntary modification (e.g. change of supplier).
Routine and periodic tests include:
![]() |
Validation of steam sterilization process is carried out according to ISO 17665-114 and guided by ISO 17665-215 and ISO 17665-316
ISO 17665-1 specifies the implementation requirements for large steam sterilizers. Technical specification ISO 17665-2 provide guidance for implementation of ISO 17665 -1. Technical specification ISO 17665-3 provide guidance to assign device in product families presenting common characteristics. These 3 texts refer to the criteria and test methods applicable to steam sterilizers complying to EN 28517.
ISO 17665-1, -2 and -3 cover hospital and industry applications of steam sterilization. Annex D of EN ISO 17665-2 refers to considerations specific to healthcare establishments.
ISO 17665-1 can also be used, by analogy, for small steam sterilizers conforming to standard EN 1306018 (sterilizers intended for loads smaller than one sterilization unit)
IQ verifies and documents the compliance of the services (water supply, steam supply if applicable, power, supporting structure, etc.) with the manufacturer’s specifications. IQ also checks the availability of documentation (Instruction for use, maintenance, calibration certificates, certificates of essential type tests) as well as statement of conformities to local, regional, or national regulations for emissions into the environment from the equipment. The installation and documentation checks may need to be repeated, in part, after risk assessment, after a relocation of the equipment or a major modification / repair to the installation.
Local regulation define safety measure and periodic inspection against the risk related to te safety of pressure vessels
OQ verifies that the equipment operates as specified by the manufacturer: OQ tests include steam quality, air leakage flow rate, thermometric test, air detector if fitted, steam penetration, load dryness, thermometric tests hollow load and Bowie-Dick test,
PQ is intended to control that reproducible results are obtained in conditions representative of real life.
Thermometric tests are performed with the most challenging load among product families and loading configurations that the sterilizing process may treat.
Advice on characterizing for difficulty should be available from the medical device manufacturer(s) and/or sterilizer manufacturer. Element to be considered when defining the most challenging load are mass (tray or container with the highest mass, the material and geometry (hollow devices), packaging (wrap, pouches, container).
Experience has shown that for a typical health care sterilization load and chamber usable space (ca. 400 l), 5 to 12 temperature sensors may be sufficient. Some sensor are placed on parts of the medical device from which is assumed (or known) that stelm penetration/air removal is most difficult to achieve.
If a process challenge device is used the challenge must be equivalent to the medical device(s).
Operational and Performance requalifications are commonly annual. The contents of the initial IQ, OQ and PQ and scheduled requalifications are predefined.
The content of the unscheduled validation is event related (for example IQ may be partially or totally repeated after a relocation of the equipment or a major repair).
Other examples of event requiring a requalification are a modification of the sterilization program or packaging A revalidation may be required (for example, modification of the pre-treatment or sterilization phase, change of the duration), after modification of a packaging. Following a load modification, revalidation can be avoided if it has been shown that the new load is equivalent to another load already validated.
Routine controls are performed at each cycle, daily, weekly or quarterly
|
in waiting for the development of ISO and EN vH2O2 standards , ISO 149371 and ISO 11138-120 remains the references for the validation of vH2O2 processes.
3 vH2O2 standards are currently in development. ISO 22441: Sterilization of health care products – Low temperature vaporized hydrogen peroxide – Requirements for the development, validation and routine control of a sterilization process for medical devices” will have the same structure as ISO 14937. Links will be made to EN 17180 “sterilizers for medical purposes – low temperature vaporized hydrogen peroxide sterilizers-requirements and testing” the future vH2O2 standard for manufacturer of H2O2 sterilizers. EN17180 will use the structure of ISO 22421(not to be mixed with ISO 22441) . ISO 11138-6: Sterilization of healthcare products – biological indicators – Part 6 – biological indicators for hydrogen peroxide sterilization processes” will confirm Geobacillus Stearothermophillus will remain the reference test microorganisms for hospital applications. Both ISO 22441 and EN 17180 drafts have agreed on the minimal variables to be measured by a vH2O2 process: Pressure, Time, Temperature and vH2O2 concentration. vH2O2 concentration can be measured by an H2O2 or deducted by other means such as pressure dynamic.
![]() |
Validation of all steps of the sterilization processes from point of use processing to aseptic presentation is key to minimize the risk of a non sterile RMD
Go to WFHSS Recommandations for process validation →
1 of 5 Validation of Steam sterilizationFor an example of validation of sterilization process
Go to Validation of steam sterilization →
For details on validations on packaging process
Go to Validation of packaging process →
3 of 5 Process validationFor generale explanation on principles of validations
Go to Principles of validations →
4 of 5 Validation of cleaning processFor details on validations on automated manual and ultrasonic assisted cleaning process
Go to Validation of cleaning process →
5 of 5“The WFHSS executive committee is pleased to put the WFHSS Guidelines at your disposal.
They have been written for all the professionals working directly or indirectly in the field of the reprocessing of Reusable Medical Devices (RMD) used in Health Care facilities.
They are the result of a consensus from a review of national practices, standards, regulations.
They are intended to provide guidance and the state of the art recommendations from an academic world society focused on science but their purpose is not to supersede local regulation, standards or guidelines.
They will be updated regularly to follow the evolution of the science and the evolution of the RMD and the technologies.
Your feedback is essential to contribute to the improvement of the Guidelines , feel free to use the form to send your comments or suggestions.
We wish you interesting reading!
On behalf of the Executive Committee
Dr Christine DENIS, President”
Combination of all technical and associated administrative actions intended to keep equipment at a state in which it can perform its required function, or restore it to such a state (ISO 11139 : 2018)
Packaging is intended to preserve the sterility of teh reusable medical device (RMD) until its use.
Rigid sterile barrier system designed to be repeatedly used (ISO 11607-1 2018)
For the puropose of these guidelines, International Standard means standardsor guidance published by international standardization organizations such as ISO or CEN.
Go to the international standards paragraph or Regulation and standards chapter
Prion is a small proteinaceous infectious unit that appear in a variety of neurodegenerative diseases, including bovine spongiform encephalopathy, Creutzfeldt-Jakob disease, and scrapie. They derive from a normal body protein that becomes irreversibly misfolded and proliferates primarily in the central nervous system.
Prions are highly resistant to disinfection and sterilization
Some sterilizing, cleaning and disinfecting formulations are known to favor the adherence and resistance of proteins, including prion on RMD surfaces.
Levels of purity of ambient air and compressed air used for drying.
Chemical, physical, biological, and radiological characteristics of water used for cleaning, disinfection, rinsing and steam sterilization of RMD.
Processing operation performed at point of use of the Reusable Medical Device (i.e. operating theatre or care unit)
Quality management includes all the activities that organizations use to direct, control, and coordinate quality. These activities include formulating a quality policy and setting quality objectives. They also include quality planning, quality control, quality assurance, and quality improvement.
In the present guide the quality management chapter includes a description of the processus approach, performance evaluation, risk and non-conformity management, documentation management and traceability
Go to quality management chapter →
The Spaulding classification qualify the RMD as non critical when they touch intact skin, semi-critical when they are brought in contact with mucous membranes and critical when they enter sterile body cavities. Processing requirements increase with level of risk involved in their use.
OHS deals with all aspects of health and safety in the workplace and has a strong focus on primary prevention of hazards (WHO : 2016)
For the present guidlelines reusable medical device (RMD) means:
The reprocessing of single use medical device is outside the scope of theses guidelines.
Medical device regulations vary between regions. Some items may not be registered as medical devices in some regions.
Sterilization is intended to renders the reusable medical device free from viable microorganisms. Sterilization is implemented on a clean RMD. Most common Sterilization process is steam. Low temperature sterilization processes are available for heat sensitive RMD
Disinfection :
Process to reduce the number of viable microorganisms to a level previously specified as being appropriate for a defined purpose (ISO 11139 : 2018)
Cleaning :
Removal of contaminants to the extent necessary for its further processing or for intended use (ISO 11139 : 2018)
Other definition: The first step required to physically remove contamination by foreign material, e.g. dust soil. It will also remove organic material such as blood, secretion, excretion and microorganisms, to prepare a medical device for sterilization or disinfection (WHO : 2016)
Cleaning may be combined to disinfection in cleaning & disinfections processes (for instance in automated washer-disinfectors)
Sterilization :
Process used to render product free from viable microorganisms (ISO 11139 : 2018)
SOP’s are written, step-by-step instructions that describe how to perform a routine activity. SOP’s aim to achieve efficiency, quality output and uniformity of performance, while reducing miscommunication and failure to comply with industry regulations.
technical operation conducted periodically to establish that the operational performance of the equipment or process remains within the limits established during validation (ISO 11139 : 2018)
Some sterilizing, cleaning and disinfecting formulations are known to favor the anchorage of proteins, including prion on RMD surfaces. For instance, dry heat, ethylene oxide, aldehyde based sterilizing and disinfecting agent such formaldehyde or glutaraldehyde, alcohol used to accelerate drying of some RMD. For this reasons, some national guidelines recommend to avoid use of this subtances or require specific precautions (for instance thorough cleaning)
“The WFHSS executive committee is pleased to put the WFHSS Guidelines at your disposal.
They have been written for all the professionals working directly or indirectly in the field of the reprocessing of Reusable Medical Devices (RMD) used in Health Care facilities.
They are the result of a consensus from a review of national practices, standards, regulations.
They are intended to provide guidance and the state of the art recommendations from an academic world society focused on science but their purpose is not to supersede local regulation, standards or guidelines.
They will be updated regularly to follow the evolution of the science and the evolution of the RMD and the technologies.
Your feedback is essential to contribute to the improvement of the Guidelines , feel free to use the form to send your comments or suggestions.
We wish you interesting reading!
On behalf of the Executive Committee
Dr Christine DENIS, President”
Non terminal sterilization is a process whereby product is not sterilized in a sterile barrier system and hence not protected from environmental and handling contamination after the sterilization cycle.
Go to Sterilization chapter →
Process whereby product is sterilized within its sterile barrier system (ISO 11139 : 2018)
Go to Sterilization chapter →
SOP’s are written, step-by-step instructions that describe how to perform a routine activity. SOP’s aim to achieve efficiency, quality output and uniformity of performance, while reducing miscommunication and failure to comply with industry regulations.
For the needs of the present guidelines outsourcing means a process in which a healthcare facility employs another organization to perform some or all reusable medical device reprocessing tasks. The organization can be another healthcare facility, a shared reprocessing unit or a private service company. The services can be perfomed within the healthcare facility or externalized.
Pyrogen are substances that induces fever. Endogenous pyrogen are low-molecular weight protein produced by phagocytic leukocytes in response to stimulation by exogenous pyrogens. Exogenous pyrogens are produced by bacterial endotoxins and other microbial product such as antigen, antibody complexes, virus
Endotoxins are lipopolysaccharide components of the cell wall of Gram-negative bacteria that are heat stable and elicits a variety of inflammatory responses in animals and humans (ISO 11139: 2018)
Ability to trace the history, application, use and location of an item (products, parts, materials, and services) or its characteristics through recorded identification data.
For the purpose of theses guidelines, training means the certified acquisition of the theoricital, practical skills and behavior adapted to the assignment. Skills are periodically controlled and updated as needed.
Routine control check that performances of process or equipment are maintained over time between 2 process validations. Routine controls can be systematic (at each cycle) or at predefined periodicities. For instance, for the sterilization process, at end of each cycle, it is checked that process parameterare are within the validated tolerances, leak tests are performed daily or at periodicity defined by local recommandations.
Go to process validation chapter →
Written indication provided by the manufacturer to ensure correct and safe use of a products (including but not limited to reusable medical device, reprocessing equipment and consumables). IFU of RMD include the instruction for reprocessing. IFU are also available for reprocessing equipment and consumables.
Identification and analyzis of potential issues that could negatively impact a given process. Risk analysis includes an evaluation of the consequence and likelihood. The issues with the more severe consequences and higher likelihood of occurence are addressed in priority. As needed measure are then to minimize the occurence or consequence. Methods are available to improve the reliability of the risk analysis.
Storage concerns mainly sterile medical device. Storage of disinfected devices may be allowed by local regulation in defined conditions.
Transport includes:
Process validation means establising, by objective evidence that a process consistently produces a results or product meeting its predeternined objective. Process validation applies to all steps of RMD reprocessing from point of use processing to storage. When equipments are used validation include their installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ).
Process validation is perfomed before implementation of a new process or equipment (initial validation). Revalidation takes place periodically (usually each year) or after event justifying total or partial revalidation (change in the process, maintenance of equipments).
Waste management groups all activities and actions required to manage waste including the collection, transport, treatment and disposal of waste together with monitoring of the waste management process. Instrument reprocessing waste are are solid (i.e., single use packaging, detergent or disinfectant empty bottles, cleaning brushes, possibly sharp devices), liquid (i.e., cleaning, disinfecting, rinsing solutions) or gazeous (i.e., sterilization, cleaning or disinfection effluent). Waste might be non hazardous, infectious, or toxic. Waste management regulation limits impact on environment and protects staff. Waste management rules are country dependent.
Traceability is the ability to trace the history, application, use and location of an item (products, parts, materials, and services) or its characteristics through recorded identification data.
Reusable Medical device (RMD)
Medical devices wich are not single use i.e. which can be reused under appropriate reprocessing conditions, for an undelimited number of time or for a predetermined number of use.