检查和装配
检查和装配 – 原则

检查和装配 – 原则

复用医疗器械 (RMD) 的检查和装配是在清洁和消毒后进行。按照规范,经过检查和装配的 RMD可以使用或进行灭菌。

在装配多部件 RMD 或 RMD 组合(例如手术包)之前和之后进行检查(即目视检查或/和技术功能测试)。

检查和装配工作站应满足工作需要(例如,操作台面和光线)。


当保养和维修的RMD返回时,需特别注意。

 

多部件 RMD 和RMD组合的检查和装配,要参考 RMD 制造商使用说明 (IFU)

手术包的装配是与诊疗部门和手术部门合作准备的。共同制定的指导书中明确:

  • 属于该手术包的器械,和每种器械的数量
  • 托盘类型(密网状、排孔状、热成型)和尺寸(足够大以允许重量和金属质量均匀分布)
  • 若需要,使用固定和保护装置(如保护尖锐易碎物品、避免运输过程中移位或震动的楔子、托盘衬垫等)
  • 为满足使用现场安全取用要求,必要时应规定RMD 的摆放位置。 

 

适用的一般规则:

  • 重器械的摆放方式不会损坏其他精密器械物品。
  • 具有较大凹面、扁平面或两者兼有的物品,可能会存水,将其侧放。
  • 对于带管腔的 RMD,管腔端口是敞开的。
  • 器械关节应张开,应备有将器械关节保持在张开位置所需附件的说明书。 

器械标识符和图片有助于简化复杂器械包的装配。

需检验:

  • 可提供RMD,RMD打包装配方式和RMD保护装置与预期使用的灭菌周期的兼容性证书
  • 灭菌剂与RMD 表面可充分接触。 尖端保护器(如果使用)允许灭菌剂穿透,松紧合适,并根据制造商的书面 IFU 使用。 仅使用经认可的附件。
  • 重量不超过职业健康和安全限制(根据每个国家/地区的指南,包括包装在内的物品的最大重量在 10 公斤至 11 公斤(25 磅)之间变化)

对于碗类物品,所有碗盆都采用同一方向摆放,并在叠放的碗盘之间放入无毛絮的吸水材料。

对于敷料,咨询产品制造商以获取针对蒸汽灭菌周期验证的包装大小、尺寸和密度的使用建议。敷料包是使用清洗过的敷料制作的。

检查,装配和质量

书面的检查标准操作流程 (SOP) 是根据质量管理原则制定的。

验证检查和装配的过程时关注:

  • SOP 是最新的,遵循 RMD 制造商的 IFU,并且符合诊疗部门和手术部门的期望
  • 执行培训并提供培训证书。
  • 职业健康和安全措施到位。
  • 追溯是可行的且有效
Screenshot-2019-09-07-at-13.22.20-e1567939504403

WFHSS 检查和装配建议

  1. 检查和装配的标准操作流程 (SOP) 遵循 RMD 制造商 IFU 并符合最终用户的期望。
  2. 检验RMD与保护装置的兼容性
  3. 检查和装配过程按照质量管理原则实施并经过过程确认
运输流程图
清洗和消毒后的RMD

Used on patient or withdrawn from packaging

Go to Control & Assembly – Principles 

1 of 5
检查和装配

RMD or multi-component RMD or surgical set

Go to Control and Assembly – Principles 

2 of 5
可用于下一步处理的RMD

Next step: use or sterilization

Go to Control and Assembly – Principles

3 of 5

SOP prpeared in partnership with user

Go to WFHSS recommendation for Control and Assembly

4 of 5

User feedback

Go to Control and Assembly and quality

5 of 5
参考文献
  1. ISO 17664-1 :医疗保健产品的处理 – 医疗器械制造商需要提供的处理医疗器械的信息 (2017)

“The WFHSS executive committee is pleased to put the WFHSS Guidelines at your disposal.
They have been written for all the professionals working directly or indirectly in the field of the reprocessing of Reusable Medical Devices (RMD) used in Health Care facilities.

They are the result of a consensus from a review of national practices, standards, regulations.
They are intended to provide guidance and the state of the art recommendations from an academic world society focused on science but their purpose is not to supersede local regulation, standards or guidelines.
They will be updated regularly to follow the evolution of the science and the evolution of the RMD and the technologies.

Your feedback is essential to contribute to the improvement of the Guidelines , feel free to use the form to send your comments or suggestions.

We wish you interesting reading!

On behalf of the Executive Committee

Dr Christine DENIS, President”  

播放视频

Maintenance

Combination of all technical and associated administrative actions intended to keep equipment at a state in which it can perform its required function, or restore it to such a state (ISO 11139 : 2018)

Packaging

Packaging is intended to preserve the sterility of teh reusable medical device (RMD) until its use.

Reusable containers

Rigid sterile barrier system designed to be repeatedly used (ISO 11607-1 2018)

International standards

For the puropose of these guidelines, International Standard means standardsor guidance published by international standardization organizations such as ISO or CEN.
Go to the international standards paragraph or Regulation and standards chapter 

Prion

Prion is a small proteinaceous infectious unit that appear in a variety of neurodegenerative diseases, including bovine spongiform encephalopathy, Creutzfeldt-Jakob disease, and scrapie. They derive from a normal body protein that becomes irreversibly misfolded and proliferates primarily in the central nervous system.
Prions are highly resistant to disinfection and sterilization

Fixative properties

Some sterilizing, cleaning and disinfecting formulations are known to favor the adherence and resistance of proteins, including prion on RMD surfaces.

Air quality

Levels of purity of ambient air and compressed air used for drying.

Water quality

Chemical, physical, biological, and radiological characteristics of water used for cleaning, disinfection, rinsing and steam sterilization of RMD.

Point of use reprocessing

Processing operation performed at point of use of the Reusable Medical Device (i.e. operating theatre or care unit)

Quality management 

Quality management includes all the activities that organizations use to direct, control, and coordinate quality. These activities include formulating a quality policy and setting quality objectives. They also include quality planning, quality control, quality assurance, and quality improvement.
In the present guide the quality management chapter includes a description of the processus approach, performance evaluation, risk and non-conformity management, documentation management and traceability

Go to quality management chapter →

Spaulding classification

The Spaulding classification qualify the RMD as non critical when they touch intact skin, semi-critical when they are brought in contact with mucous membranes and critical when they enter sterile body cavities. Processing requirements increase with level of risk involved in their use.

Occupational health and safety (OHS)

OHS deals with all aspects of health and safety in the workplace and has a strong focus on primary prevention of hazards (WHO : 2016)

Reusable medical device (RMD)

For the present guidlelines reusable medical device (RMD) means:

  • an item that is not intended by its manufacturer for single use.
  • A medical device designated or inrtended by the manufacturer as suitable for reprocessing.

The reprocessing of single use medical device is outside the scope of theses guidelines.
Medical device regulations vary between regions. Some items may not be registered as medical devices in some regions.

Sterilization

Sterilization is intended to renders the reusable medical device free from viable microorganisms. Sterilization is implemented on a clean RMD. Most common Sterilization process is steam. Low temperature sterilization processes are available for heat sensitive RMD

Disinfection :

Process to reduce the number of viable microorganisms to a level previously specified as being appropriate for a defined purpose (ISO 11139 : 2018) 

Cleaning :

Removal of contaminants to the extent necessary for its further processing or for intended use (ISO 11139 : 2018)  

Other definition: The first step required to physically remove contamination by foreign material, e.g. dust soil. It will also remove organic material such as blood, secretion, excretion and microorganisms, to prepare a medical device for sterilization or disinfection (WHO : 2016) 

Cleaning may be combined to disinfection in cleaning & disinfections processes (for instance in automated washer-disinfectors)

Tooltip Content

Sterilization :

Process used to render product free from viable microorganisms (ISO 11139 : 2018)

  • courthouse-3-1 Sterilization is intended to renders the reusable medical device (RMD) free from viable microorganisms.

Standard operating procedures

SOP’s are written, step-by-step instructions that describe how to perform a routine activity. SOP’s aim to achieve efficiency, quality output and uniformity of performance, while reducing miscommunication and failure to comply with industry regulations.

Routine control (or routine test)

technical operation conducted periodically to establish that the operational performance of the equipment or process remains within the limits established during validation (ISO 11139 : 2018) 

Fixative properties

Some sterilizing, cleaning and disinfecting formulations are known to favor the anchorage of proteins, including prion on RMD surfaces. For instance, dry heat, ethylene oxide, aldehyde based sterilizing and disinfecting agent such formaldehyde or glutaraldehyde, alcohol used to accelerate drying of some RMD. For this reasons, some national guidelines recommend to avoid use of this subtances or require specific precautions (for instance thorough cleaning)

“The WFHSS executive committee is pleased to put the WFHSS Guidelines at your disposal.
They have been written for all the professionals working directly or indirectly in the field of the reprocessing of Reusable Medical Devices (RMD) used in Health Care facilities.

They are the result of a consensus from a review of national practices, standards, regulations.
They are intended to provide guidance and the state of the art recommendations from an academic world society focused on science but their purpose is not to supersede local regulation, standards or guidelines.
They will be updated regularly to follow the evolution of the science and the evolution of the RMD and the technologies.

Your feedback is essential to contribute to the improvement of the Guidelines , feel free to use the form to send your comments or suggestions.

We wish you interesting reading!

On behalf of the Executive Committee

Dr Christine DENIS, President” 

Non terminal sterilization

Non terminal sterilization is a process whereby product is not sterilized in a sterile barrier system and hence not protected from environmental and handling contamination after the sterilization cycle.

Terminal sterilization

Process whereby product is sterilized within its sterile barrier system (ISO 11139 : 2018)

Standard operating procedures (SOP's)

SOP’s are written, step-by-step instructions that describe how to perform a routine activity. SOP’s aim to achieve efficiency, quality output and uniformity of performance, while reducing miscommunication and failure to comply with industry regulations.

Outsourcing

For the needs of the present guidelines outsourcing means a process in which a healthcare facility employs another organization to perform some or all reusable medical device reprocessing tasks. The organization can be another healthcare facility, a shared reprocessing unit or a private service company. The services can be perfomed within the healthcare facility or externalized.

Pyrogen

Pyrogen are substances that induces fever. Endogenous pyrogen are low-molecular weight protein produced by phagocytic leukocytes in response to stimulation by exogenous pyrogens. Exogenous pyrogens are produced by bacterial endotoxins and other microbial product such as antigen, antibody complexes, virus

Endotoxin

Endotoxins are lipopolysaccharide components of the cell wall of Gram-negative bacteria that are heat stable and elicits a variety of inflammatory responses in animals and humans (ISO 11139: 2018)

Traceability

Ability to trace the history, application, use and location of an item (products, parts, materials, and services) or its characteristics through recorded identification data.

Training

For the purpose of theses guidelines, training means the certified acquisition of the theoricital, practical skills and behavior adapted to the assignment. Skills are periodically controlled and updated as needed.

Routine controls

Routine control check that performances of process or equipment are maintained over time between 2 process validations. Routine controls can be systematic (at each cycle) or at predefined periodicities. For instance, for the sterilization process, at end of each cycle, it is checked that process parameterare are within the validated tolerances, leak tests are performed daily or at periodicity defined by local recommandations. 

Go to process validation chapter →

Instruction for use (IFU)

Written indication provided by the manufacturer to ensure correct and safe use of a products (including but not limited to reusable medical device, reprocessing equipment and consumables). IFU of RMD include the instruction for reprocessing. IFU are also available for reprocessing equipment and consumables.   

Risk analysis

Identification and analyzis of potential issues that could negatively impact a given process. Risk analysis includes an evaluation of the consequence and likelihood. The issues with the more severe consequences and higher likelihood of occurence are addressed in priority. As needed measure are then to minimize the occurence or consequence. Methods are available to improve the reliability of the risk analysis. 

Storage

Storage concerns mainly sterile medical device. Storage of disinfected devices may be allowed by local regulation in defined conditions.

Transport

Transport includes:

  • The transfer of soiled reusable medical device from the point of use to the reprocessing site and
  • the transfer of the sterilized or disinfected RMD to the point of use.

Process validation

Process validation means establising, by objective evidence that a process consistently produces a results or product meeting its predeternined objective. Process validation applies to all steps of RMD reprocessing from point of use processing to storage. When equipments are used validation include their installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ).
Process validation is perfomed before implementation of a new process or equipment (initial validation). Revalidation takes place periodically (usually each year) or after event justifying total or partial revalidation (change in the process, maintenance of equipments).

Waste management

Waste management groups all activities and actions required to manage waste including the collection, transport, treatment and disposal of waste together with monitoring of the waste management process. Instrument reprocessing waste are are solid (i.e., single use packaging, detergent or disinfectant empty bottles, cleaning brushes, possibly sharp devices), liquid (i.e., cleaning, disinfecting, rinsing solutions) or gazeous (i.e., sterilization, cleaning or disinfection effluent). Waste might be non hazardous, infectious, or toxic. Waste management regulation limits impact on environment and protects staff. Waste management rules are country dependent.

Traceability

Traceability is the ability to trace the history, application, use and location of an item (products, parts, materials, and services) or its characteristics through recorded identification data.

Reusable Medical device (RMD) 

Medical devices wich are not single use i.e. which can be reused under appropriate reprocessing conditions, for an undelimited number of time or for a predetermined number of use.