Click words or phrases color for definitions and descriptions
When a word or phrase is the title of a chapter, a link is provided to give direct access to the chapter
The texts governing medical devices reprocessing and education activities may be grouped in 3 categories :
In case of overlaps the applicable regulation always supersedes standards and guidelines.
ISO sterilization and washer-disinfector standards are under the responsibility of Technical Committee 198 (TC 198). EN disinfection standards are under CEN TC 102, and CEN TC 204.
Standards are prepared by experts nominated by their national delegations.
Proposal for a new standard may be submitted via a national delegation or by ISO or CEN technical committees. Once the scope of new work item proposal is approved by a majority of ISO or EN country members, the project is assigned by the Technical Committee to a working group. The first draft is often prepared by the initiators of the project. It is circulated for comments by national committees. Comments are then reviewed one by one by the working group experts. Several rounds of drafts, votes and comments review are usually needed until a consensual text is obtained and submitted for final approval. Along the road, collaborative tests may be needed to align on performance thresholds. Details can be obtained on ISO or CEN websites. Once published, Standards are periodically submitted for revision, amendment, confirmation or withdrawal. |
Figure 1 illustrates ISO and CE framework for sterilization activities. It is provided for illustration purpose only. This diagram is subject to constant change and only valid for the indicated period. Official list and status are displayed on ISO and CEN websites
![]() |
![]() |
Some standards are intended to all stakeholders involved in device reprocessing i.e. healthcare facilities, manufacturers of reprocessing equipement or consumables, as well as manufacturers of medical devices and service providers.
This is the case of the ISO sterilization process standards for steam 3,4,5 , low temperature hydrogen peroxide6, ethylene oxide7, low temperature steam formaldehyde8 , dry heat9.
ISO 1493710 applies to sterilization technologies which do not have a dedicated standard.
For example, the sterilization process standards3,4,5,6,7,8,9,10 will be used by:
|
Packaging standards 11,12 and guidance 13 will be used by healthcare facilities for the packaging process validation. Most of the requirements included in the standards will however be for manufacturers of single use devices and for manufacturer of sterile barrier systems (SBS) and for packaging materials.
Other standards or guidances are primarily intended for manufacturers of sterilizing equipment, consumables or medical devices. They define the requirements and test methods to be applied to justify compliance. Those standards are however also of interest for healthcare facilities. For example, steriizers standards may be used to prepare tenders,
Examples of non TC 198 standards to which ISO and EN sterilization standards commonly refer are:
ISO 1099351,52,53 for the evaluation of the biological safety of the material after exposure to a sterilization or cleaning process |
ISO standards for thermal and chemical washer-disinfectors35,36,37,38,39 are also primarily intended for manufacturers. However they contain guidance for healthcare facilities to check the conformity of installed washer-disinfector throughout its working life. ISO 15883-440 and EN 1644241 are dedicated to heat sensitive flexible endoscopes (reprocessing and storage respectively). ISO 15883-542 describes test method and criteria for cleaning efficacy
Other ISO TC 198 standards and guidance displayed on the above diagram come in support of other TC 198 documents.
Examples of non TC 198 standards to which ISO and EN sterilization standards commonly refer are:
ISO 1099351,52,53 for the evaluation of the biological safety of the material after exposure to a sterilization or cleaning process |
In the above diagram key standards not belonging to ISO TC 198, EN TC 102 or TC 204 but to which ISO and EN standards refer, are also displayed
Examples of non TC 198 standards to which ISO and EN sterilization standards commonly refer are:
ISO 1099351,52,53 for the evaluation of the biological safety of the material after exposure to a sterilization or cleaning process |
The following diagram illustrates how the various sterilization, washer-disinfector, BI, CI, processing information and packaging standard may be used by healthcare facilities, manufacturers of processing equipments and consumables as well as medical device manufacturers
Normative language can sometimes be perceived as difficult to translate into practical considerations for healthcare facilities. Some possible explanations are listed below (not exhaustive):
The sterilization process standards contain a foreword, an introduction,12 normative paragraphs and a variable number of normative or informative annexes.
The first three normative paragraphs equally apply to all categories of potential user of the standards.
Maintaining process effectiveness the healthcare facility may contract with the sterilizer manufacturer for maintenance, operational requalification, performance requalification or qualified third party for performance requalification |
Unlike for sterilization, there are no universal ISO Standards for evaluation of disinfection chemicals. Depending on the region different methods and acceptance criteria apply.
Disinfectant are commonly classified in 3 categories: low level, intermediate level and high level. However the borders between each category are not universally defined. For example high level disinfection is always associated to some level of efficacy on spores but the efficacy threshold on spores is not the same.
An example of a commonly used definition for a sporicidal formulation is: a germicide that inactivates all microbial pathogens, except large numbers of bacterial endospores (Rutala 1990, Spaulding 1970). However the applicable test method and criteria is country dependent different. For example :
|
Unlike for sterilization in health care facilities, where sterility test are performed only with the germs predetermined as the most resistant, disinfection claims are for specific categories of germs : bacteria, fungi and yeast, virus (envelopped and non envelopped) and mycobacteria and spores, expressed in log reduction, with log reduction objective which may differ between catgories of microorganismes and region.
In the European community the test method to be applied for the various type of disinfection claims are defined by EN 14885.
EN 1488561 covers human medicine, veterinary, food, industrial, domestic and institutional areas. Within the medical domain EN 14885 proposes adapted substantiating strategies for various field of application hand hygien (hygienic handrub, hygienic handwash, Surgical Handrub or-wash), surface disinfection (with or with mechanical action and airbone), textile disinfection and aqueous system and instrument disinfection. The rest of this paragraph will focus on disinfection of instruments. EN 14885 list test in suspension (phase 2.1) or on carrier (Phase 2.2). For medical devices, phase 2.2 carrier tests are more representative of real life and closer to AOAC and ASTM approach. Interference matter is added to simulates soil (in clean or dirty condition depending on the application) Test are perfomed at the concentration and exposure time specified by the manufacturer. At the end of the exposure time a neutralizer is added to stop the microbiocidal activity before counting to check that the targeted log reduction is achieved.
|
Unlike for disinfectant there are no international standard for the evaluation of cleaning efficacy but the efficacy of the cleaning process (i.e. combination of mechanical and chemical action) can be evaluated according to ISO 15883-5.
The illustration below is an attempt to summarize the place of terminal sterilization, liquid sterilization and the various categories of disinfectants.
![]() |
Guidelines may be specific to a given topic, for example steam sterilization or global like WFHSS guidelines. Some guidelines are in free access like WFHSS. For other, fees must be paid.
WFHSS guidelines and other worlwide or regional guidance do not supersede local guidelines when they exist. See Guidelines for example of international, local or regional guidelines
It is often considered that significant differences exists between guidelines. In reality, core reprocessing directives are common and differences are in the how.
“The WFHSS executive committee is pleased to put the WFHSS Guidelines at your disposal.
They have been written for all the professionals working directly or indirectly in the field of the reprocessing of Reusable Medical Devices (RMD) used in Health Care facilities.
They are the result of a consensus from a review of national practices, standards, regulations.
They are intended to provide guidance and the state of the art recommendations from an academic world society focused on science but their purpose is not to supersede local regulation, standards or guidelines.
They will be updated regularly to follow the evolution of the science and the evolution of the RMD and the technologies.
Your feedback is essential to contribute to the improvement of the Guidelines , feel free to use the form to send your comments or suggestions.
We wish you interesting reading!
On behalf of the Executive Committee
Dr Christine DENIS, President”
“The WFHSS executive committee is pleased to put the WFHSS Guidelines at your disposal.
They have been written for all the professionals working directly or indirectly in the field of the reprocessing of Reusable Medical Devices (RMD) used in Health Care facilities.
They are the result of a consensus from a review of national practices, standards, regulations.
They are intended to provide guidance and the state of the art recommendations from an academic world society focused on science but their purpose is not to supersede local regulation, standards or guidelines.
They will be updated regularly to follow the evolution of the science and the evolution of the RMD and the technologies.
Your feedback is essential to contribute to the improvement of the Guidelines , feel free to use the form to send your comments or suggestions.
We wish you interesting reading!
On behalf of the Executive Committee
Dr Christine DENIS, President”
For the present guidlelines reusable medical device (RMD) means:
The reprocessing of single use medical device is outside the scope of theses guidelines.
Medical device regulations vary between regions. Some items may not be registered as medical devices in some regions.
Sterilization is intended to renders the reusable medical device free from viable microorganisms. Sterilization is implemented on a clean RMD. Most common Sterilization process is steam. Low temperature sterilization processes are available for heat sensitive RMD
Disinfection :
Process to reduce the number of viable microorganisms to a level previously specified as being appropriate for a defined purpose (ISO 11139 : 2018)
Cleaning :
Removal of contaminants to the extent necessary for its further processing or for intended use (ISO 11139 : 2018)
Other definition: The first step required to physically remove contamination by foreign material, e.g. dust soil. It will also remove organic material such as blood, secretion, excretion and microorganisms, to prepare a medical device for sterilization or disinfection (WHO : 2016)
Cleaning may be combined to disinfection in cleaning & disinfections processes (for instance in automated washer-disinfectors)
Sterilization :
Process used to render product free from viable microorganisms (ISO 11139 : 2018)
Reusable Medical device (RMD)
Medical devices wich are not single use i.e. which can be reused under appropriate reprocessing conditions, for an undelimited number of time or for a predetermined number of use.